A-SERIES A625 LIFT (2017-09-15)
- Starting date:
- September 15, 2017
- Posting date:
- October 5, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- Hospitals, Healthcare Professionals, General Public
- Identification number:
- RA-64750
Affected products
A-SERIES A625 LIFT
Reason
During internal cycle testing of A-Series lifts Prism Medical Ltd. Has found that the A-Series Lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). They have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.
Affected products
A-SERIES A625 LIFT
Lot or serial number
Not Applicable
Model or catalog number
324118