This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Sequencer

Starting date:
August 22, 2011
Posting date:
September 17, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-15534

Recalled Products

  1. Sequencer

Reason

Issues may cause Mosaiq to fail to recognize the designated decimal separator as valid, resulting in misinterpretation and/or incorrect display of fractional numeric data, or to correctly interpret date values under very particular conditions.

Affected products

A. Sequencer

Lot or serial number
  • Mosaiq Version 1.60V1 to 2.30.01
Model or catalog number
  • 10-0020
Companies
Manufacturer

Elekta Business Area Software Systems Impac Medical Systems Inc.

13723 Riverport Drive, Suite 100

Maryland Heights

63043

Missouri

UNITED STATES