This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Sequencer
- Starting date:
- August 22, 2011
- Posting date:
- September 17, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-15534
Recalled Products
- Sequencer
Reason
Issues may cause Mosaiq to fail to recognize the designated decimal separator as valid, resulting in misinterpretation and/or incorrect display of fractional numeric data, or to correctly interpret date values under very particular conditions.
Affected products
A. Sequencer
Lot or serial number
- Mosaiq Version 1.60V1 to 2.30.01
Model or catalog number
- 10-0020
Companies
- Manufacturer
-
Elekta Business Area Software Systems Impac Medical Systems Inc.
13723 Riverport Drive, Suite 100
Maryland Heights
63043
Missouri
UNITED STATES