Health product recall

SEMA 3

Brand(s)
Last updated

Summary

Product
SEMA 3
Issue
Medical devices - Device compatibility
What to do

Contact the manufacturer if you require additional information.

Audience
Industry

Affected products

Affected products Lot or serial number Model or catalogue number
SEMA 3 Not applicable. 6020.009411
SEMA 3 Not applicable. 6020.012099
SEMA 3 Not applicable. 6020.009412

Issue

When an ECG is aquired with DEFIGARD Touch 7 or ARGUS PRO LifeCare 3 with the newest software, there might be a discrepancy between the aquiring device and SEMA. This results from inadequate processing by SEMA to an already filtered signal. SEMA should be updated to version 25.10. 

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Schiller Ag

Altgasse 68, Baar, Zug, Switzerland, 6341

Published by
Health Canada
Audience
Industry
Recall class
Type II
Recall date
Identification number
RA-82336

Get notified

Receive emails about new and updated recall and safety alerts.