SEMA 3
Brand(s)
Last updated
Summary
Product
SEMA 3
Issue
Medical devices - Device compatibility
What to do
Contact the manufacturer if you require additional information.
Audience
Industry
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| SEMA 3 | Not applicable. | 6020.009411 |
| SEMA 3 | Not applicable. | 6020.012099 |
| SEMA 3 | Not applicable. | 6020.009412 |
Issue
When an ECG is aquired with DEFIGARD Touch 7 or ARGUS PRO LifeCare 3 with the newest software, there might be a discrepancy between the aquiring device and SEMA. This results from inadequate processing by SEMA to an already filtered signal. SEMA should be updated to version 25.10.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Schiller Ag
Altgasse 68, Baar, Zug, Switzerland, 6341
Published by
Health Canada
Audience
Industry
Recall class
Type II
Recall date
Identification number
RA-82336
Get notified
Receive emails about new and updated recall and safety alerts.