This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Screwdriver Shaft, Hexagonal, Large

Starting date:
February 28, 2017
Posting date:
March 24, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62778

Reason

Depuy Synthes is initiating a medical device removal of one lot of the screwdriver shaft, hexagonal, large, diameter 3.5 mm, length 165 mm, for quick coupling, because of a complaint that the location slot for the screw holding sleeve was machined 3 mm out of position.

Affected products

Screwdriver Shaft, Hexagonal, Large

Lot or serial number

7722501

Model or catalog number

314.560

Companies

Manufacturer

SYNTHES GMBH

EIMATTSTRASSE 3

OBERDORF

4436

SWITZERLAND