Health product recall

Schiller EASY PULSE®

Brand(s)
Last updated

Summary

Product
Schiller EASY PULSE®
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

EASY PULSE®

4000.000701

3.940410

EASY PULSE®

4000.000728

3.940410

EASY PULSE®

4000.000729

3.940410

EASY PULSE®

4000.000701

3.940409

EASY PULSE®

4000.000728

3.940409

EASY PULSE®

4000.000729

3.940409

Issue

The root cause has been identified as excessive counterforce applied to the piston at start-up, which prevents the piston from initiating movement. With the release of software version 1.1.3, the initial acceleration of the piston was reduced. However, this adjustment has inadvertently increased the device's sensitivity to counterforces, resulting in premature activation of the safety stop under certain conditions.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Schiller Ag

Altgasse 68, Baar, Zug, Switzerland, 6341

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-78405

Get notified