Sapphire Multi-Therapy Infusion Pump (2013-10-08)

Starting date:
October 8, 2013
Posting date:
November 13, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-36621

Recalled Products

Sapphire Multi-Therapy Infusion Pump

Reason

Under certain cumulative conditions the time spent in each of the steps may deviate from the programmed time for the step. The infusion time will therefore be longer or shorter than the programmed time and the amount delivered at each step will also differ from the programmed amounts, but the total volume of the infusion will be the same as programmed.

Affected products

Sapphire Multi-Therapy Infusion Pump

Lot or serial number

All serial numbers

Model or catalog number
  • 17000-028-0020
  • 17000-028-0021
Companies
Manufacturer

Q Core Medical Ltd.

29 Yad Haruzim St., P.O. Box 8639,

Netanya

4250529

ISRAEL