Salmonella Antiserum H:t
- Starting date:
- January 17, 2017
- Posting date:
- February 9, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-62128
Reason
A decrease in antibody level was detected in this lot which means that agglutination with salmonella strains having weak expression of H:g,m,s,t and H:f,g,t appears after the definition of a positive reaction (10 sec), and therefore would be conceived as negative when in fact it is a positive.
Affected products
Salmonella Antiserum H:t
Lot or serial number
696H-H6
Model or catalog number
40336
Companies
- Manufacturer
-
SSI DIAGNOSTICA A/S
2 HERREDSVEJEN
HILLROD
3400
DENMARK