This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Salmonella Antiserum H:t

Starting date:
January 17, 2017
Posting date:
February 9, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62128

Reason

A decrease in antibody level was detected in this lot which means that agglutination with salmonella strains having weak expression of H:g,m,s,t and H:f,g,t appears after the definition of a positive reaction (10 sec), and therefore would be conceived as negative when in fact it is a positive.

Affected products

Salmonella Antiserum H:t

Lot or serial number

696H-H6

Model or catalog number

40336

Companies

Manufacturer

SSI DIAGNOSTICA A/S

2 HERREDSVEJEN

HILLROD

3400

DENMARK