Health product recall

Salem Sump Dual Lumen Stomach Tubes (2018-11-05)

Starting date:
November 5, 2018
Posting date:
November 28, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68418

Affected products

Salem Sump Dual Lumen Stomach Tubes

Reason

Cardinal is recalling the Salem Sump Dual Lumen Stomach Tubes due to incorrect packaging. In the affected lot of Salem Sump Dual Lumen Stomach Tubes, the 10fr product was packaged in a 16fr labeled pouch. The use of a 10 fr Salem Sump Tube when a 16fr is needed could result in decreased fluid collection, with the potential to affect patient care, particularly in an emergency situation where the correct size is not readily available. Cardinal health is not aware of any reports of injury or harm related to this issue.

Affected products

Salem Sump Dual Lumen Stomach Tubes

Lot or serial number
  • 1812115564
Model or catalog number
  • 8888264911
Companies
Manufacturer
Cardinal Health
Patient Recovery Business from Covidien LLC
15 Hampshire Street
Mansfield
02048
UNITED STATES