S5 Single Roller Pump 150 and S5 Mast Roller Pump 150 and 85
- Starting date:
- September 20, 2012
- Posting date:
- October 15, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public (GP), Healthcare Professionals, Hospitals
- Identification number:
- RA-15175
Recalled Products:
- S5 Single Roller Pump 150
- S5 Mast Roller Pump 150 and 85
Reason
The manufacturer recently received reports of a "Fault in Motor Controller" alarm message. If this occurs, the pump stops and audible alarm sounds and a visual alarm is displayed. The "Fault in Motor Controller" message is displayed on the system panel. On some occasions the circuit breaker of the pump on the emergency power pack may also trip.
Affected products
A. S5 Single Roller Pump 150
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
10-80-00
Companies
- Manufacturer
-
Sorin Group Deutschland GMBH
B. S5 Mast Roller Pump 150 and 85
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- 10-88-00
- 10-88-60
Companies
- Manufacturer
-
Sorin Group Deutschland GMBH