Health product recall

S-Series Equipment Management Syetem (2021-02-02)

Starting date:
February 2, 2021
Posting date:
February 19, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-75011



Last updated:
2021-02-19

Affected Products 

S-Series Equipment Management Syetem

Reason

Stryker has become aware that there is a likelihood that the S-Series Equipment Management System Capacitive Touch Controls on the MFRs rotational electrical magnetic brakes may disengage with the use of an electrocautery device. The potential hazard associated is the boom may experience unintentional horizontal drifting with the use of an electrocautery device that could lead to patient injury needing medical/surgical intervention.

Affected products

S-Series Equipment Management Syetem

Lot or serial number

2258046-2863490

Model or catalog number

0682-400-550

Companies
Manufacturer

Stryker Communications Inc

571 Silveron Blvd

Flower Mound

75028

Texas

UNITED STATES