Health product recall

RX Daytona System Lipase and Lipase (LPS) (2018-02-28)

Starting date:
February 28, 2018
Posting date:
March 21, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66200

Affected products

A. RX Daytona System - Lipase

B. LIPASE (LPS)

Reason

Randox have now released further steps for contamination avoidance with the lipase assay on RX instruments. Users should refer to the instrument testing order specifically arranging chemistries so that lipase and triglycerides are the last two chemistries in the test running order.