Rotor-Gene Q Software (2020-02-03)
- Starting date:
- February 3, 2020
- Posting date:
- April 15, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72735
Last updated: 2020-04-15
Affected Products
Rotor-Gene Q Software
Reason
RGQ SW version 2.3.4 completing a lims export, the .CSV file reports the calculated concentration result value as a logarithmic value. For quantitative analysis there is a risk that results will be significantly under the expected value and reflected in the lims system could lead to a false negative result.
Affected products
Rotor-Gene Q Software
Lot or serial number
Not applicable.
Model or catalog number
9019101
Companies
- Manufacturer
-
QIAGEN GmbH
QIAGEN Strasse 1
Hilden
40724
GERMANY