Health product recall

Rotor-Gene Q Software (2020-02-03)



Starting date:
February 3, 2020
Posting date:
April 15, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72735


Last updated:

Affected Products

Rotor-Gene Q Software

Reason

RGQ SW version 2.3.4 completing a lims export, the .CSV file reports the calculated concentration result value as a logarithmic value. For quantitative analysis there is a risk that results will be significantly under the expected value and reflected in the lims system could lead to a false negative result.

Affected products

Rotor-Gene Q Software

Lot or serial number

Not applicable.

Model or catalog number

9019101

Companies
Manufacturer

QIAGEN GmbH

QIAGEN Strasse 1

Hilden

40724

GERMANY