Health product recall

Rotating IV Pole (2017-12-12)

Starting date:
December 12, 2017
Posting date:
December 21, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Hospitals, General Public, Healthcare Professionals
Identification number:
RA-65504

Affected products

Rotating IV Pole

Reason

GE Healthcare has recently become aware of the potential for the Rotating IV Pole to fall due to excessive tap drill depth resulting in a thin wall section of the block bracket assembly which is the bracket that attaches the pole to the bed. A fall could result in an injury to a bystander holding the patient or to the operator. There have been no injuries as a result of this issue.

Affected products

Rotating IV Pole

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

6600-0851-800

Companies
Manufacturer
Ohmeda Medical, A Division of Datex Ohmeda Inc.
8880 GORMAN ROAD
LAUREL
20723
Maryland