This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Rotating IV Pole (2017-12-12)

Starting date:
December 12, 2017
Posting date:
December 21, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Hospitals, General Public, Healthcare Professionals
Identification number:
RA-65504

Affected products

Rotating IV Pole

Reason

GE Healthcare has recently become aware of the potential for the Rotating IV Pole to fall due to excessive tap drill depth resulting in a thin wall section of the block bracket assembly which is the bracket that attaches the pole to the bed. A fall could result in an injury to a bystander holding the patient or to the operator. There have been no injuries as a result of this issue.

Affected products

Rotating IV Pole

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

6600-0851-800

Companies
Manufacturer

Ohmeda Medical, A Division of Datex Ohmeda Inc.

8880 GORMAN ROAD

LAUREL

20723

Maryland

UNITED STATES