Health product recall

Roche Diagnostics Assays (2018-11-18)

Starting date:
November 18, 2018
Posting date:
January 11, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68744

Affected products

  1. COBAS Integra Lactate Gen 2 (LACT2)
  2. IGM-2 (TINA-QUANT IGM GEN.2) FOR USE ON C701 ANALYZER
  3. IGM-2 (TINA-QUANT IGM GEN.2) ROCHE/HITACHI 904/911/912/917 AN MOD. P ANALY.
  4. IGM-2 (TINA-QUANT IGM GEN.2) FOR USE ON COBAS C SYSTEMS ANALYZERS
  5. LACT2 (LACTATE GEN.2) FOR COBAS C701 ANALYZER
  6. BILIRUBIN TOTAL GEN.3 (BILT3)

Reason

During internal study, selected endogenous drug interferences in serum/plasma, urine and cerebrospinal fluid were assessed on all Roche tests and new endogenous interferences claims for three tests (BIL-T Gen.3, IGM Gen2 and LACT Gen.2) were found.

Affected products

A. COBAS Integra Lactate Gen 2 (LACT2)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

03183700190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY


B. IGM-2 (TINA-QUANT IGM GEN.2) FOR USE ON C701 ANALYZER

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

05220726190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY


C. IGM-2 (TINA-QUANT IGM GEN.2) ROCHE/HITACHI 904/911/912/917 AN MOD. P ANALY.

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

03507041190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY


D. IGM-2 (TINA-QUANT IGM GEN.2) FOR USE ON COBAS C SYSTEMS ANALYZERS

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

03507190190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY


E. LACT2 (LACTATE GEN.2) FOR COBAS C701 ANALYZER

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

05171881190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY


F. BILIRUBIN TOTAL GEN.3 (BILT3)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

05795320190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY