This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

RIWO Net Mobile System Trolley

Starting date:
March 19, 2013
Posting date:
June 17, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-34181

Recalled products

  1. RIWO Net Mobile System Trolley

Reason

In one single case so far the VESA mounting plate broke at its mounting point while the user was adjusting the inclination (Tilt). Internal investigations have shown that as a results of a manufacturing error the weld of the mounting plate at the mounting point of the rotary joint was faulty. There is therefore a risk of a monitor fastened to the VESA mounting plate failling down when the tilt is adjusted. The VESA mounting plate is part of the device RIWO Net mobile system trolley.

Affected products

A. RIWO Net Mobile System Trolley

Lot or serial number

All Serial Numbers 

Model or catalog number

Catalogue Number:

  • 32115.011

Part Numbers:

  • 321114452
  • 321114453
  • 64330914
Companies
Manufacturer

Richard Wolf GmbH

Pforzheimer Str.32, Postfach 1164/1165,

Knittlingen, Baden-Wuerttemberg

75438

GERMANY