RIWO Net Mobile System Trolley
- Starting date:
- March 19, 2013
- Posting date:
- June 17, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-34181
Recalled products
- RIWO Net Mobile System Trolley
Reason
In one single case so far the VESA mounting plate broke at its mounting point while the user was adjusting the inclination (Tilt). Internal investigations have shown that as a results of a manufacturing error the weld of the mounting plate at the mounting point of the rotary joint was faulty. There is therefore a risk of a monitor fastened to the VESA mounting plate failling down when the tilt is adjusted. The VESA mounting plate is part of the device RIWO Net mobile system trolley.
Affected products
A. RIWO Net Mobile System Trolley
Lot or serial number
All Serial Numbers
Model or catalog number
Catalogue Number:
- 32115.011
Part Numbers:
- 321114452
- 321114453
- 64330914
Companies
- Manufacturer
-
Richard Wolf GmbH
Pforzheimer Str.32, Postfach 1164/1165,
Knittlingen, Baden-Wuerttemberg
75438
GERMANY