REVOLUTION CT (2016-05-12)

Starting date:
May 12, 2016
Posting date:
May 30, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58616

Affected products 

A. REVOLUTION CT

Reason

GE Healthcare has become aware of a potential safety issue in the Revolution CT patient table that can result in unintended motion in cases involving large patients. The Revolution CT cradle can descend on its own after activation by the user. This can put the patient at risk for harm involving the potential impinging or crushing of a hand if caught between the cradle and CT inner gantry bore. No injuries have been reported to date related to this issue.

Affected products

A. REVOLUTION CT

Lot or serial number

More than 10 numbers, contact manufacturer. 

Model or catalog number

5590000

Companies
Manufacturer
GE Medical Systems LLC
3200 North Grandview Blvd
Waukesha
53188
Wisconsin
UNITED STATES