Reveal Linq Mobile Manager
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or Serial Number |
Model or catalog number |
|---|---|---|
Reveal Linq Mobile Manager |
RLA223716G |
MSW002 |
Issue
The Reveal Linq Mobile Manager application (MSW002) versions 2.01, 2.02, 2.03.0, and 2.03.01. These versions of LMM have an updated insertion workflow that, if not followed properly, increases the likelihood the patient connector is removed from the device by users before the Reveal Linq device is programmed successfully.
This incomplete programming can result in the device's inability to detect and record some or all the following episode types: AT, AF, Tachy, Asystole, Brady, and patient-activated symptom episodes.
Recall start date: Mar 22, 2022
Additional information
Details
Medtronic Inc.
8200 Coral Sea Street N.E., Mounds View, Minnesota
United States, 55112