This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

ResMed Astral (2015-05-08)

Starting date:
May 8, 2015
Posting date:
July 9, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54152

Recalled Product

  1. Astral 150
  2. Astral 100

Reason

One incident reported in Germany involving circuit disconnection of a patient in Hospital where the device alarms did not operate because all alarms had been disengaged.

Affected products

A. Astral 150

Lot or serial number

N/A

Model or catalog number
  • 27003
  • 27073
Companies
Manufacturer
Resmed Limited
1 Elizabeth Macarthur Drive
Bella Vista
2153
AUSTRALIA

B. Astral 100

Lot or serial number

N/A

Model or catalog number

27001

Companies
Manufacturer
Resmed Limited
1 Elizabeth Macarthur Drive
Bella Vista
2153
AUSTRALIA