Health product recall

Reprocessed LASSO 2515 Nav ECO Variable Catheter (2018-12-16)

Starting date:
December 16, 2018
Posting date:
January 9, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68668

Affected products

Reprocessed LASSO 2515 Nav ECO Variable Catheter

Reason

Stryker received a report of failure mode. A trend was identified by manufacturer indicating that an EEPROM chip error code may occur when Reprocessed 2515 NAV eco Variable Electrophysiology (EP) Catheters are used with CARTO EP Navigation Systems. There have been no injuries reported as a result of this event.

Affected products

Reprocessed LASSO 2515 Nav ECO Variable Catheter

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number
  • D134301
Companies
Manufacturer
Stryker Sustainability Solutions, Inc.
1810 West Drake Dr.
Tempe
85283
Arizona
UNITED STATES