Renasys Edge Canister With Solidifier
Brand(s)
Last updated
Summary
Product
Renasys Edge Canister With Solidifier
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Renasys Edge Canister With Solidifier(300,800ml) | More than 10 numbers, contact manufacturer. | 66803139 66803140 |
Issue
Smith & Nephew Medical Ltd. has initiated a field action to voluntarily remove several lots of Renasys Edge Canisters due to black particulate matter from the carbon filter component of the canisters entering the npwt pump and potentially causing damage to the pump.
Recall start date: March 9, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
| Smith & Nephew Medical Ltd. |
| 101 Hessle Road, Hull, Great Britain, HU3 2BN |
Published by
Health Canada
Audience
General public
Recall class
Type II
Identification number
RA-75306
Get notified