Health product recall

REMISOL ADVANCE (2021-07-15)



Starting date:
July 15, 2021
Posting date:
July 29, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-76101


Last updated:

Affected Products 

REMISOL ADVANCE

Reason

Beckman Coulter discovered an issue with the DiffPad feature of Remisol Advance versions 2.0 And 2.1. Whenever keyboard settings are configured to numerical values there is a chance that additional cell counts are added while performing a manual differential

Affected products

REMISOL ADVANCE

Lot or serial number

C44703AD

Model or catalog number

VERSION 2.0

VERSION 2.1

Companies
Manufacturer

Normand - Info Sasu

7 rue Frederic Degeorge

Arras Cedex

62028

FRANCE