Health product recall

REMISOL Advance

Last updated

Summary

Product
REMISOL Advance
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

REMISOL Advance

All lots.

VERSION 1.12
VERSION 1.2
VERSION 1.3
VERSION 1.5
VERSION 2.1
VERSION 1.1
VERSION 2.0
VERSION 1.8
VERSION 1.4
VERSION 2.2
VERSION 1.11
VERSION 1.6
VERSION 1.7
VERSION 1.9
VERSION 1.10
VERSION 2.3

Issue

Beckman Coulter has identified an issue with REMISOL Advance when receiving a sample order from the Laboratory Information System (LIS) if the patient's Date of Birth (DOB) is not entered or is unknown:

  • If the LIS sends 01/01/1900 as the DOB, REMISOL Advance using the ASTM host driver incorrectly interprets the age as 0 days.
  • If the LIS sends an empty DOB, REMISOL Advance using the HL7 host driver incorrectly interprets the age as 0 days.

Recall start date: April 7, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Normand - Info Sasu

7 Rue Frederic Degeorge, Arras Cedex, France, 62028

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77344

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe