Reliance Vision Multi-Chamber Washer Disinfector
Last updated
Summary
Product
Reliance Vision Multi-Chamber Washer Disinfector
Issue
Medical devices - Performance issue
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Industry
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Reliance Vision Multi-Chamber Washer Disinfector | All lots. | FH92-072 FH91-072 FH82-072 FH81-072 FH82-042 |
Issue
Steris has identified that in the remote occurrence in which the electrical contactor component present in the drying chambers of the reliance vision multi-chamber washer/disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.
Recall start date: March 29, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Corporation Steris Canada (Also Operating As Steris Canada ULC)
490 Boul. Armand-Paris, Quebec, Quebec, Canada, G1C 8A3
Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-73570
Get notified