Reliance Vision Multi-Chamber Washer Disinfector
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Reliance Vision Multi-Chamber Washer Disinfector |
All lots. |
FH92-072 |
Issue
Steris has identified that in the remote occurrence in which the electrical contactor component present in the drying chambers of the reliance vision multi-chamber washer/disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.
Recall start date: March 29, 2023
Additional information
Details
Corporation Steris Canada (Also Operating As Steris Canada ULC)
490 Boul. Armand-Paris, Quebec, Quebec, Canada, G1C 8A3
Get notified
Receive notifications for new and updated recalls and alerts by category.