Reliance Synergy Washer/Disinfector
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or Serial Number |
Model or catalog number |
|---|---|---|
Reliance Synergy Washer/Disinfector |
More than 10 numbers, contact manufacturer. |
FH03-052, FH03-072, FH03-053, FH03-082, FH03-083 |
Issue
Steris has identified that in the remote occurrence in which the electrical contactor component present in the drying chamber of the Reliance Synergy Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire. This could result in smoke inhalation or a burn which may require medical intervention.
Recall start date: Apr 14, 2022
Additional information
Details
Steris Canada ULC
490 Boul. Armand-Paris, Quebec, Quebec
Canada, G1C 8A3
Get notified
Receive emails about new and updated recall and safety alerts.