This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Reliance 130 and 130L Cart Washer

Starting date:
July 14, 2012
Posting date:
September 3, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-23819

Recalled Products

A. Reliance 130 Cart Washer

B. Reliance 130L Cart Washer

Reason

The Reliance 130 and 130L Cart Washer/Disinfector is a high capacity mechanical washer intended for use in the washing, low-level disinfecting, and drying of carts, utensils, beds and other miscellaneous reusable items used in the care of patients at healthcare facilities.  In review of the reported events, Steris has learned users are not always familiar with the intended function of the door open/close commands and the washer's safety features.  To ensure users are properly operating the Reliance 130 and 130L Cart Washer a "door close" confirmation button and door open delay will be installed on the Cart Washer(s).

Affected products

A. Reliance 130 Cart Washer

Lot or serial number

3606499005

Model or catalog number

FW02-XXX

Companies
Manufacturer

Steris Canada Corporation



B. Reliance 130L Cart Washer

Lot or serial number

3628809011

Model or catalog number

FW01-XXX

Companies
Manufacturer

Steris Canada Corporation