Health product recall

RegenKit® BCT-3

Brand(s)
Last updated

Summary

Product
RegenKit® BCT-3
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
RegenKit® BCT-3 303 301 308 309 BCT-3

Issue

Clients have reported several cases of patients having inflammatory reactions after injection with platelet-rich plasma (PRP), characterized by joint effusion and/or pain. Particles in the sodium citrate solution have been observed, with an irregular appearance of the separation gel and the presence of a white layer on the gel surface. To clarify, the separation gel is used for blood separation. It creates a physical barrier between the red blood cells and the PRP. Only the plasma is reinjected into the patient; the gel is not.

Recall start date: August 2, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Regen Lab SA

En Budron B2, Le Mont, Switzerland, 1052

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72394