RAYSTATION (2016-07-01)
- Starting date:
- July 1, 2016
- Posting date:
- July 19, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- Hospitals, General Public, Healthcare Professionals
- Identification number:
- RA-59408
Affected products
A. RAYSTATION
Reason
This issue concerns an issue with the evaluation of biological clinical goals in RayStation 3.0. RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7 and RayStation 5. Biological clinical goals for an adapted plan based on another planning CT than the original planning CT will show incorrect values.
Affected products
A. RAYSTATION
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- "4.0.3
- 4.7
Companies
- Manufacturer
-
Raysearch Laboratories AB,
Sveavagen 25,
Stockholm
11134
SWEDEN