Health product recall

RAYSTATION (2016-02-12)

Starting date:
February 12, 2016
Posting date:
March 10, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57468

Affected Products

  1. RAYSTATION

Reason

Issue found with some of the tools used for drawing in and interacting with the patient views and beam's eye views. The issue concerns all tools that are used with the left mouse button held down. These tools can become unsynchronized with the stored data if other commands such as for example a right mouse click are used at the same time as the left mouse button is held down.

Affected products

  1. RAYSTATION

Lot or serial number
  • 3.0
  • 3.5
  • 4.0
  • 4.5
  • 4.7
Model or catalog number
  • 4.0.3
  • 4.7