This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Raystation

Starting date:
January 18, 2017
Posting date:
February 14, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62190

Reason

Issue found with accelerated computation on Nvidia GPU (graphic processing unit) in Raystation 4.5, Raystation 4.7, Raystation 5 and Raystation 6 when used in a non-standard Citrix Environment where the GPU settings are configured with a virtual "Citrix" graphics board (graphics card).

Affected products

Raystation

Lot or serial number

  • 4.5.0.19
  • 4.5.1.14
  • 4.5.2.7
  • 4.7.0.15
  • 4.7.1.10
  • 4.7.2.5
  • 4.7.3.13
  • 4.7.4.4
  • 4.7.5.4
  • 5.0.0.37
  • 5.0.1.11
  • 5.0.2.35

Model or catalog number

  • 4.0.3
  • 5

Companies

Manufacturer

Raysearch Laboratories AB,

Sveavagen 25,

Stockholm

11134