Raystation
- Starting date:
- January 18, 2017
- Posting date:
- February 14, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-62190
Reason
Issue found with accelerated computation on Nvidia GPU (graphic processing unit) in Raystation 4.5, Raystation 4.7, Raystation 5 and Raystation 6 when used in a non-standard Citrix Environment where the GPU settings are configured with a virtual "Citrix" graphics board (graphics card).
Affected products
Raystation
Lot or serial number
- 4.5.0.19
- 4.5.1.14
- 4.5.2.7
- 4.7.0.15
- 4.7.1.10
- 4.7.2.5
- 4.7.3.13
- 4.7.4.4
- 4.7.5.4
- 5.0.0.37
- 5.0.1.11
- 5.0.2.35
Model or catalog number
- 4.0.3
- 5
Companies
- Manufacturer
-
Raysearch Laboratories AB,
Sveavagen 25,
Stockholm
11134
SWEDEN