This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

RAYSTATION

Starting date:
September 1, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60358

Affected Products

Reason

When a region of interest (ROI) with material override overlaps a fixation or support ROI, the material used for dose calculation in the overlapping region may not be as intended. Also, when a support or fixation ROI overlaps the patient outline (external ROI) in a dose grid voxel along the outer surface of the patient outline, the density used for dose calculation in that voxel may not be as intended.

Affected products

RAYSTATION

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

2.5

3.0

4.0

4.5

4.7

Companies
Manufacturer

Raysearch Laboratories AB,

Sveavagen 25,

Stockholm

11134