This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Rapidpoint 405 System Measurement Cartridges

Starting date:
September 7, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60278

Affected Products

Rapidpoint 405 System Measurement Cartridges

Reason

Siemens has confirmed that there is an error in the value assignment for the Chloride Calibrators in Rapidpoint 400 and Rapidpoint 405 measurement cartridge serial numbers 2610900214 through 2621001325.

Affected products

Rapidpoint 405 System Measurement Cartridges

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

10283222

130522

130523

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New York

UNITED STATES