RADREX-I (2015-12-21)
- Starting date:
- December 21, 2015
- Posting date:
- January 22, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-56740
Affected Products
- RADREX-I
Reason
When an operator performs radiography with Wireless FPD, the following message may be displayed on the monitor: "The image cannot be acquired from FPD. Press ok to try retrieving." however, when the operator selects "ok" or "cancel" on the message window, the image is not acquired and is lost.
Affected products
A. RADREX-I
Lot or serial number
S/N: W8B14X2014; S/N: W8B1512017
Model or catalog number
DRAD-3000E_/W8
Companies
- Manufacturer
-
Toshiba Medical Systems Corporation
1385 Shimoishigami
OTAWARA-SHI
324-8550
JAPAN