Health product recall

RadPRO Mobile 3 40kW (2018-11-02)

Starting date:
November 2, 2018
Posting date:
November 28, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68434

Affected products

RadPRO Mobile 3 40kW

Reason

There is the potential for the four (4) retention screw brass tabs, which secure the system's collimator to the tube head mount, to wear out which could result in the system's collimator separating from the tube head mount and potentially falling off of the system.

Affected products

RadPRO Mobile 3 40kW

Lot or serial number
  • G65175
  • G65382
  • G65384
  • G65418
  • G67250
  • G68470
  • G70521
  • G70524
  • G70913
  • G73624
Model or catalog number
  • MOBILE 3 40KW 50G/C
Companies
Manufacturer

Virtual Imaging, Inc.

720 SOUTH POWERLINE ROAD, SUITE E

DEERFIELD BEACH

33442

Florida

UNITED STATES