RadPRO Mobile 3 40kW (2018-11-02)
- Starting date:
- November 2, 2018
- Posting date:
- November 28, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68434
Affected products
RadPRO Mobile 3 40kW
Reason
There is the potential for the four (4) retention screw brass tabs, which secure the system's collimator to the tube head mount, to wear out which could result in the system's collimator separating from the tube head mount and potentially falling off of the system.
Affected products
RadPRO Mobile 3 40kW
Lot or serial number
- G65175
- G65382
- G65384
- G65418
- G67250
- G68470
- G70521
- G70524
- G70913
- G73624
Model or catalog number
- MOBILE 3 40KW 50G/C
Companies
- Manufacturer
-
Virtual Imaging, Inc.
720 SOUTH POWERLINE ROAD, SUITE E
DEERFIELD BEACH
33442
Florida
UNITED STATES