This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

RadPRO 40kW - Digital Mobile X-Ray System (2015-07-13)

Starting date:
July 13, 2015
Posting date:
October 22, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55490

Affected Products

RadPRO 40kW - Digital Mobile X-Ray System

Reason

Electrostatic energy may be stored in systems that can cause a short-circuit which, in turn, can generate an electromagnetic pulse which might disturb the electronics of the digital motion control board producing uncontrolled movements and such uncontrolled movements could cause the system to bump into the user or patient.

Affected products

RadPRO 40kW - Digital Mobile X-Ray System

Lot or serial number

S/N between G29393 and G62480

Model or catalog number

MOBILE 3 40KW 50G/C

Companies
Manufacturer

Virtual Imaging, Inc.

720 SOUTH POWERLINE ROAD, SUITE E

DEERFIELD BEACH

33442

Florida

UNITED STATES