RadPRO 40kW - Digital Mobile X-Ray System (2015-07-13)

Starting date:
July 13, 2015
Posting date:
October 14, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55334

Affected Products

RadPRO 40kW - Digital Mobile X-Ray System

Reason

Electrostatic energy may be stored in systems that can cause a short-circuit which, in turn, can generate an electromagnetic pulse which might disturb the electronics of the digital motion control board producing uncontrolled movements. Such uncontrolled movements could cause the system to bump into the user or patient.

Affected products

RadPRO 40kW - Digital Mobile X-Ray System

Lot or serial number

Serial number between G29393 and G62480

Model or catalog number

MOBILE 3 40KW 50G/C