Radiant Warmer and Birthing Room Warmer
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Radiant Warmer |
RW82-1C |
MU20503 |
Birthing Room Warmer |
WBR82-1 |
MU20512 |
Issue
During the course of global market surveillance activities, the manufaturer has become aware of cases in which the resuscitaire infant radiant warmer with the optional scale could display inaccurate weight values. There is a potential for the mattress cover material to slightly interfere with the corner brackets and/or side panels of the bassinet. When this condition is present, the scale may not "zero" correctly during a "weigh" function, which may lead to an inaccurate weight measurement of the infant.
Recall start date: April 28, 2023
Additional information
Details
Draeger Medical Systems, Inc.
3135 Quarry Road, Telford, Pennsylvania, United States, 18969
Get notified
Receive emails about new and updated recall and safety alerts.