Health product recall

QuickDraw 22 & 25 Venous Cannula (2019-12-17)



Starting date:
December 17, 2019
Posting date:
January 10, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72075


Last updated:

Affected Products

  1. QuickDraw 22 Venous Cannula
  2. QuickDraw 25 Venous Cannula

Reason

Edwards Lifesciences has identified a potential safety risk that may occur during the use of QuickDraw Venous Cannula models QD22 and QD25 when used in an unintended manner. The QuickDraw Cannula is not designed, tested, or intended for use beyond 6 hours. When the cannula is used for longer than 6 hours, particularly during ECMO procedures and the associated patient movement during ECMO, there is a risk that separation of the cannula from its connector can occur; which can result in significant blood loss.

Affected products

  1. QuickDraw 22 Venous Cannula

Lot or serial number

All lots since Nov 14/17

Model or catalog number

QD22

Companies
Manufacturer

Edwards Lifesciences LLC

1 Edwards Way

Irvine

92614

California

UNITED STATES



B. QuickDraw 25 Venous Cannula

Lot or serial number

All lots since Nov 14/17

Model or catalog number

QD25

Companies
Manufacturer

Edwards Lifesciences LLC

1 Edwards Way

Irvine

92614

California

UNITED STATES