This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Qube Compact Monitor (2014-06-06)

Starting date:
June 6, 2014
Posting date:
June 24, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40183

Recalled products

  1. Qube Compact Monitor 

Reason

Potential issue with Qube compact monitors that are "docking  station ready" - that have had a port cover plate removed allowing the monitor to be used with the optional Qube docking station. If a docking station ready Qube monitor is powered up undocked, connected to the network via an ethernet hardwire connection, and then disconnected from the ethernet hardwire connection and inserted into a Qube docking station, a switch condition will cause the monitor to lose the network connection. All bedside alarms will continue to function normally. No one has been injured as a result of this issue.

Affected products

A. Qube Compact Monitor

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 91390
Companies
Manufacturer

Spacelabs Healthcare Inc.

35301 SE Center Street

Snoqualmie

98065

Washington

UNITED STATES