Health product recall

qUAntify Control and qUAntify Plus Control (2018-06-18)

Starting date:
June 18, 2018
Posting date:
June 29, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67142

Affected products

  1. qUAntify Control
  2. qUAntify Plus Control

Reason

Bio-rad has found that values for blood and ketones can gradually decrease, while pH values can gradually increase over the course of the lifespan of these product lots.

Affected products

A. qUAntify Control

Lot or serial number
  • 78770
  • 78780
Model or catalog number
  • 975
  • 975X
Companies
Manufacturer

Bio-Rad Laboratories

9500 Jeronimo Road

Irvine

92618

California

UNITED STATES



B. qUAntify Plus Control

Lot or serial number
  • 79610
  • 79611
  • 79612
  • 79620
  • 79621
  • 79622
  • 79630
  • 79631
  • 79632
  • 79640
  • 79641
  • 79642
  • 79650
  • 79651
  • 79652
  • 79660
  • 79662
  • 80510
  • 80520
  • 80530
  • 80540
  • 80550
  • 80560
Model or catalog number
  • 962
  • 962X
  • 963
  • 964
  • 995
  • 995X
Companies
Manufacturer

Bio-Rad Laboratories

9500 Jeronimo Road

Irvine

92618

California

UNITED STATES