Health product recall

Quadrox-I Adult Oxygenator W/Integrated Hardshell Reservoir

Last updated

Summary

Product
Quadrox-I Adult Oxygenator W/Integrated Hardshell Reservoir
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Quadrox-I Adult Oxygenator W/Integrated Hardshell Reservoir 3000358089 VKMO 70000

Issue

Maquet Cardiopulmonary's monthly inspection of oxygenators for the presence of the Softline coating revealed that two out of ten oxygenators tested did not pass. The Softline coating is a vital component in the biocompatibility of blood contact components used in extracorporeal circulation. This means that the oxygenators installed in the products listed above do not have the Softline coating. The issue is limited to one lot number in Canada.

Recall Start Date: August 20, 2024 

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Maquet Cardiopulmonary Gmbh

Kehler Strasse 31, Rastatt, Germany, 76437

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75990

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe