This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

QIAsymphony DSP Virus/Pathogen Midi Kit (2014-12-08)

Starting date:
December 8, 2014
Posting date:
December 29, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43057

Recalled Products   

QIAsymphony DSP Virus/Pathogen Midi Kit 

Reason

Some reagent cartridges of this kit lot contain reagent troughs that are not filled correctly with buffer QSW2. As a consequence, the nucleic acids purified with affected reagent cartridges do not meet the expected quality due to contamination with inhibitory substances. The performance of the resulting nucleic acids in downstream applications is decreased.

Affected products

QIAsymphony DSP Virus/Pathogen Midi Kit 

Lot or serial number

148028163

Model or catalog number

937055

Companies
Manufacturer

Qiagen GmbH,

Qiagen Strasse 1,

Hilden

40724

GERMANY