QIAcube Connect MDX
Brand(s)
Last updated
Summary
Product
QIAcube Connect MDX
Issue
Medical devices - Software defect
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or serial number |
Model or catalog number |
---|---|---|
QIAcube Connect MDX |
S/N: 30011 |
9003070 |
Issue
The information detailed on the screen when the info icon is clicked is incorrect and should state that the volume is exact, not a minimal volume. This could cause a user to pipette an incorrect volume which could lead to mistaken results or sample loss. The main screen and user manual are correct but this supplemental information screen could confuse a user. Discovered internally at Qiagen.
Recall start date: 2021-11-18
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices
Companies
Qiagen Gmbh |
Qiagen Strasse 1, Hilden
Germany, 40724
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63672
Get notified