Health product recall

QIAcube Connect MDX

Brand(s)
Last updated

Summary

Product
QIAcube Connect MDX
Issue
Medical devices - Software defect
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected Products

Lot or serial number

Model or catalog number

QIAcube Connect MDX

S/N: 30011
S/N: 30036

9003070

Issue

The information detailed on the screen when the info icon is clicked is incorrect and should state that the volume is exact, not a minimal volume. This could cause a user to pipette an incorrect volume which could lead to mistaken results or sample loss.  The main screen and user manual are correct but this supplemental information screen could confuse a user. Discovered internally at Qiagen.

Recall start date: 2021-11-18

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices
Companies
Qiagen Gmbh

Qiagen Strasse 1, Hilden

 Germany, 40724

Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63672

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe