Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products

Lot or serial number

Model or catalog number

Pushtracker E2 & E3 Utilized With Smartdrive MX2+ Power Assist Device

1.0.02
1.0.04
1.0.01
1.0.03
1.0.00

PUSHTRACKER E3
PUSHTRACKER E2

Issue

Through design and development work, Max Mobility has identified a software issue with the application. When multiple processes are running on the watch's central processing unit, the application may crash unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.

Recall start date: May 2, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Physical medicine
Companies

Max Mobility LLC

300 Duke Drive, Lebanon, Tennessee, United States, 37090

Published by
Health Canada
Audience
General public
Industry
Recall class
Type II
Identification number
RA-73692