Health product recall

Pulmonary Vein Ablation Catheter Gold (2019-03-03)

Starting date:
March 3, 2019
Posting date:
March 22, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69382

Affected Products

Pulmonary Vein Ablation Catheter Gold

Reason

As per Post Market Surveillance Report (PMSR) for PVAC GOLD/PVAC/MASC/MAAC/T-VAC Multi-Array Ablation Catheters, 10092437DOC Rev 1D: Phrenic Nerve Injury (PNI) exhibited higher occurrence of harm than predicted. PNI is a known possible complication for atrial fibrillation ablation. So, assessment to update the IFU to add warnings and precautions for PNI was made.

Affected products

Pulmonary Vein Ablation Catheter Gold

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

990078

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES