Provido, Proveo 8 and M530 OHX Microscopes
Brand(s)
Last updated
Summary
Product
Provido, Proveo 8 and M530 OHX Microscopes
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected Products | Lot or serial number | Model or catalog number |
---|---|---|
M530 OHX – Optics Carrier | 10448737 | 291122001 11221001 281122001 140223001 60422002 140222002 301122002 230822001 |
Leica Proveo 8 Ophthalmic Microscope System | 10449058 | More than 10 numbers, contact manufacturer. |
Provido | 10448976 10448950 | 110722001 130722001 140722001 |
Issue
Complaint trending revealed an adverse trend of complaints stating that a LMS surgical operating microscope unexpectedly shut down caused by a defective 24VDC power supply which is used inside the surgical operating microscopes Provido, Proveo 8 and M530 OHX. Based on the results of the investigations there is a reasonable probability that the MOSFET in the power supply unit of these microscopes will overheat and become defective.
Recall Start Date: August 30, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices
Health products - Medical devices - Ear, nose and throat
Health products - Medical devices - General and plastic surgery
Companies
Leica Microsystems (Schweiz) Ag
Max Schmidheiny-Strasse 201, Heerbrugg, St. Gallen, Switzerland, 9435
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74338
Get notified