Proteus XR/a General Purpose Radiographic System
- Starting date:
- February 5, 2015
- Posting date:
- March 6, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-43813
Recalled products
- Proteus XR/a General Purpose Radiographic System
Reason
GE Healthcare has recently become aware of a bucky device having restricted vertical movement which results in a breakage of its supporting cable. A fall of a bucky while the system is in use could result in an injury to a patient or operator.
Affected products
A. Proteus XR/a General Purpose Radiographic System
Lot or serial number
All lots
Model or catalog number
- 2259988
Companies
- Manufacturer
-
GE Hualun Medical Systems Co. Ltd.