Protease 3 (2020-01-23)
- Starting date:
- January 23, 2020
- Posting date:
- February 7, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72263
Last updated: 2020-02-07
Affected Products
Protease 3
Reason
The data button, also called registration button, indicates an incorrect expiration date of 30-Nov-2020, while the box barrel labels capture the correct expiration date of 30-Nov-2019. If you rely on the system software to track the expiration of the kit, it would allow the reagent to be used past its real expiration.
Affected products
Protease 3
Lot or serial number
E12261
Model or catalog number
5266718001
760-2020
Companies
- Manufacturer
-
Ventana Medical Systems Inc.
1910 E Innovation Park Drive,
Tucson
85755
Arizona
UNITED STATES