Progressa (Frame)
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Progressa (Frame) |
X206AW8012 |
P7500A000158 |
Issue
During verification testing, it was discovered that a subset of the beds under test were experiencing bent compression links when raising the head section and when maximum patient weight was simulated. This failure mode was due to a change that was made to how the compression links attach to the bed frame. This may cause the compression links to bend under normal use with heavy (500 lbs) patients.
Recall start date: October 25, 2022
Additional information
Details
Hill-Rom, Inc.
1069 State Route 46 East, Batesville, Indiana, United States, 47006-9167
Get notified