Health product recall

Progressa (Frame)

Brand(s)
Last updated

Summary

Product
Progressa (Frame)
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected products

Lot or serial number

Model or catalog number

Progressa (Frame)

X206AW8012
X159AW7380
X206AW8010
X206AW8011
X159AW7378
X206AW8020
X159AW7375

P7500A000158
P7500A000842

Issue

During verification testing, it was discovered that a subset of the beds under test were experiencing bent compression links when raising the head section and when maximum patient weight was simulated. This failure mode was due to a change that was made to how the compression links attach to the bed frame. This may cause the compression links to bend under normal use with heavy (500 lbs) patients.

Recall start date: October 25, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies

Hill-Rom, Inc.

1069 State Route 46 East, Batesville, Indiana, United States, 47006-9167

Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-72179

Get notified

Receive emails about new and updated recall and safety alerts.