This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Progressa Bed Mattress (2014-03-19)

Starting date:
March 19, 2014
Posting date:
April 14, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39009

Recalled Products

Progressa Bed Mattress

Reason

Hill-Rom has become aware that certain Progressa beds with integrated mattresses may have a potential issue during turn assist mode. An error code 31C9 may be displayed on the graphical user interface and the wrench light will turn on indicating service required. The 31C9 error means the turn assist bladder is unable to hold the required pressure to operate as intended.

Affected products

Progressa Bed Mattress

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

P7520A19

P7520A21

P7520A23

P7520A3

Companies
Manufacturer

Hill-Rom Company, Inc.

1069 State Route 46 East

Batesville

47006-9167

Indiana

UNITED STATES