Health product recall

ProGrade (2018-01-16)

Starting date:
January 16, 2018
Posting date:
January 29, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65840

Affected products

ProGrade

Reason

Philips has become aware that if the wifi connection between the Skyplate detector and HP transfer point is weak, an image may fail to transfer from the Skyplate detector to the system. The image remains in the memory of the detector, but cannot be transferred wirelessly or by use of the backup cable. To continue, the operator can reset the Skyplate detector by removing its batteries, but the acquired image is lost, and a re-take is necessary.

Affected products

ProGrade

Lot or serial number

Not applicable

Model or catalog number

9897-010-01041

Companies
Manufacturer

Philips Medical Systems DMC GmbH

Rontgenstrabe 24

Hamburg

22335

GERMANY