ProGrade (2018-01-16)
- Starting date:
- January 16, 2018
- Posting date:
- January 29, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-65840
Affected products
ProGrade
Reason
Philips has become aware that if the wifi connection between the Skyplate detector and HP transfer point is weak, an image may fail to transfer from the Skyplate detector to the system. The image remains in the memory of the detector, but cannot be transferred wirelessly or by use of the backup cable. To continue, the operator can reset the Skyplate detector by removing its batteries, but the acquired image is lost, and a re-take is necessary.
Affected products
ProGrade
Lot or serial number
Not applicable
Model or catalog number
9897-010-01041
Companies
- Manufacturer
-
Philips Medical Systems DMC GmbH
Rontgenstrabe 24
Hamburg
22335
GERMANY